Comparison of standard and derived 12-lead electrocardiograms for diagnosis of coronary angioplasty-induced myocardial ischemia

      This paper is only available as a PDF. To read, Please Download here.

      Abstract

      To determine whether a derived 12-lead electrocardiogram (ECG) would demonstrate typical ST-segment changes of ischemia during percutaneous transluminal coronary angioplasty (PTCA), 207 patients were monitored with continuous 12-lead ST-segment monitoring during angioplasty. Additionally, to compare the derived and standard ECGs during known periods of ischemia with PTCA balloon inflation, 151 patients were recorded with both electrocardiographic methods during the procedure. Of the 207 patients recorded with the derived ECG, 171 (83%) had typical ischemic ST-segment changes during PTCA balloon inflation. The amplitudes of these ST deviations were similar to those observed during transient myocardial ischemia observed in clinical settings (median peak ST deviation, 225 μV). There was agreement regarding presence or absence of ischemia in 150 of the 151 patients recorded with both derived and standard electrocardiographic methods (>99% agreement). With use of the standard ECG as the “gold standard” for ischemia diagnosis, there were no false-positive results and only 1 false-negative result with the derived ECG. Furthermore, there was nearly perfect agreement between the two 12-lead methods in terms of anterior versus inferior wall patterns of ischemia. Future studies are required to determine whether continuous monitoring with a derived ECG would improve diagnosis and lead to better patient outcomes.
      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to American Journal of Cardiology
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Krucoff M
        Identification of high-risk patients with silent myocardial ischemia after percutaneous transluminal coronary angioplasty by multilead monitoring.
        Am J Cardiol. 1988; 61: 29F-34F
        • Krucoff MW
        • Parente AR
        • Bottner RK
        • Renzi RH
        • Stark KS
        • Shugoll RA
        • Ahmed SW
        • DeMichele J
        • Stroming SL
        • Kent KM
        Stability of multilead ST-segment “fingerprints” over time after percutaneous transluminal coronary angioplasty and its usefulness in detecting reocclusion.
        Am J Cardiol. 1988; 61: 1232-1237
        • Bush HS
        • Ferguson JJ
        • Angelini P
        • Willerson JT
        Twelve-lead electrocardiographic evaluation of ischemia during percutaneous transluminal coronary angioplasty and its correlation with acute reocclusion.
        Am Heart J. 1991; 121: 1591-1599
        • MacFarlane RW
        Lead systems.
        in: Comprehensive Electrocardiology. vol. 1. Pergamon Press, New York1989: 315-352
        • Pipberger HV
        • Bialek SM
        • Perloff JK
        • Schnaper HW
        Correlation of clinical information in the standard 12-lead ECG and in a corrected orthogonal 3-lead ECG.
        Am Heart J. 1961; 61: 34-43
        • Dower GE
        EASI 12-Lead Electrocardiography.
        Totemite Publishers, Point Roberts, WA1996
        • Dower GE
        • Yakush A
        • Nazzal SB
        • Jutzy RV
        • Ruiz CE
        Deriving the 12-lead electrocardiogram from four (EASI) electrodes.
        J Electrocardiol. 1988; : S182-S187
        • Frank E
        An accurate, clinically practical system for spatial vectorcardiography.
        Circulation. 1956; 13: 737-749
        • Drew BJ
        • Adams MG
        • Pelter MM
        • Wung SF
        ST segment monitoring with a derived 12-lead electrocardiogram is superior to routine cardiac care unit monitoring.
        Am J Critical Care. 1996; 5: 198-206
        • Dower GE
        • Feild D
        Progress in design of 5-electrode (EASI) lead system (abstr).
        J Electrocardiol. 1997; (in press)
        • Krucoff MW
        • Wagner NB
        • Pope JE
        • Mortara DM
        • Jackson YR
        • Bottner RK
        • Wagner GS
        • Kent KM
        The portable programmable microprocessor-driven real-time 12-lead electrocardiographic monitor: a preliminary report of a new device for the noninvasive detection of successful reperfusion or silent coronary reocclusion.
        Am J Cardiol. 1990; 65: 143-148
        • Drew BJ
        • Scheinman MM
        • Evans GT
        Comparison of a vectorcardiographically derived 12-lead electrocardiogram with the conventional electrocardiogram during wide QRS complex tachycardia, and its potential application for continuous bedside monitoring.
        Am J Cardiol. 1992; 69: 612-618