American Journal of Cardiology
Volume 105, Issue 1 , Pages 95-99 , 1 January 2010

Time-Dependent Risk of Fidelis Lead Failure

  • Image Result

    Cumulative hazard of Sprint Fidelis implantable defibrillator lead failure (solid line) ± 95% CIs (dashed lines).

    Cumulative hazard of Sprint Fidelis implantable defibrillator lead failure (solid line) ± 95% CIs (dashed lines).

  • Image Result

    Log time plotted against log cumulative hazard (slope 2.13, 95% CI 1.98 to 2.27, p <0.001).

    Log time plotted against log cumulative hazard (slope 2.13, 95% CI 1.98 to 2.27, p <0.001).

  • Image Result

    Conditional survival probabilities for Sprint Fidelis implantable defibrillator leads that are functioning normally 1 year, 2 years, and 3 years after implantation.

    Conditional survival probabilities for Sprint Fidelis implantable defibrillator leads that are functioning normally 1 year, 2 years, and 3 years after implantation.

  • Image Result

    Cumulative probability of Sprint Fidelis implantable defibrillator lead failure based on LVEF groups (<20%, 20% to 45%, >45%).

    Cumulative probability of Sprint Fidelis implantable defibrillator lead failure based on LVEF groups (<20%, 20% to 45%, >45%).

PII: S0002-9149(09)02220-6

doi: 10.1016/j.amjcard.2009.08.655

American Journal of Cardiology
Volume 105, Issue 1 , Pages 95-99 , 1 January 2010