Comparison of Dual-Energy Computed Tomography of the Heart With Single Photon Emission Computed Tomography for Assessment of Coronary Artery Stenosis and of the Myocardial Blood Supply

  • Balazs Ruzsics, MD, PhD

      Affiliations

    • Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, South Carolina
  • ,
  • Florian Schwarz, MD

      Affiliations

    • Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, South Carolina
  • ,
  • U. Joseph Schoepf, MD

      Affiliations

    • Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, South Carolina
    • Division of Cardiology, Department of Medicine, Medical University of South Carolina, Charleston, South Carolina
    • Corresponding Author InformationCorresponding author: Tel: (843) 876-7146; fax: (843) 876-3157
  • ,
  • Yeong Shyan Lee, MD

      Affiliations

    • Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, South Carolina
    • Department of Diagnostic Radiology, Tan Tock Seng Hospital, Singapore
  • ,
  • Gorka Bastarrika, MD, PhD

      Affiliations

    • Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, South Carolina
    • Department of Radiology, University of Navarra, Pamplona, Spain
  • ,
  • Salvatore A. Chiaramida, MD

      Affiliations

    • Division of Cardiology, Department of Medicine, Medical University of South Carolina, Charleston, South Carolina
  • ,
  • Philip Costello, MD

      Affiliations

    • Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, South Carolina
  • ,
  • Peter L. Zwerner, MD

      Affiliations

    • Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, South Carolina
    • Division of Cardiology, Department of Medicine, Medical University of South Carolina, Charleston, South Carolina

published online 05 June 2009.

To evaluate the performance of dual-energy computed tomography (CT) for integrative imaging of the coronary artery morphology and the myocardial blood supply, 36 patients (15 women, mean age 57 ± 11 years) with equivocal or incongruous single photon emission CT (SPECT) results were investigated by a single-contrast medium-enhanced, retrospectively electrocardiographic-gated dual-energy CT (DECT) scan with simultaneous acquisition of high and low x-ray spectra. Thirteen patients subsequently underwent invasive coronary angiography (ICA). The DECT data were used to reconstruct anatomic coronary CT angiographic images and to map the myocardial iodine distribution within the left ventricular myocardium. Two independent observers analyzed all DECT studies for stenosis and myocardial iodine defects. A segmental comparison was performed between the stress/rest SPECT perfusion defects and DECT iodine defects and between the ICA and coronary CT angiographic findings for stenosis. The sensitivity, specificity, positive predictive value, negative predictive value, and accuracy were estimated, along with the κ statistics. Overall, DECT had 92% sensitivity and 93% specificity, with 93% accuracy for detecting any type of myocardial perfusion defect seen on SPECT. Contrast defects at DECT correctly identified 85 (96%) of 89 fixed and 60 (88%) of 68 reversible myocardial perfusion defects. The interobserver agreement was very good (weighted κ = 0.87). Compared with ICA, coronary CT angiography had 90% sensitivity, 94% specificity, and 93% accuracy for the detection of >50% stenosis. In conclusion, our initial experience suggests that DECT, as a single examination, might be promising for the integrative analysis of the coronary artery morphology and the myocardial blood supply and is in good agreement with ICA and SPECT.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Dr. Schoepf is a medical consultant for Bayer-Schering, Wayne, New Jersey, Bracco, Princeton, New Jersey, General Electric, Princeton, New Jersey, Medrad, Indianola, Pennsylvania, Siemens, Malvern, Pennsylvania, and TeraRecon, San Mateo, California and receives research support from Bayer-Schering, Bracco, General Electric, Medrad, and Siemens; Dr. Costello is a medical consultant for Bracco and receives research support from Siemens; and Dr. Zwerner receives research support from Boehringer-Ingelheim, Bristol Myers Squib, Bracco, and Siemens.

 This work was supported in part by the Hounsfield Award of the Society of Computed Body Tomography and Magnetic Resonance (SCBT-MR).

PII: S0002-9149(09)00813-3

doi:10.1016/j.amjcard.2009.03.051

American Journal of Cardiology
Volume 104, Issue 3 , Pages 318-326, 1 August 2009